Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Astro-Med ASC-902: (1 year) CAL VER /5500,34401 DATE: 26-Aug-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.6 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 9 NUMBER OF LINES: 111 CONFIGURATION: HP 34401A CONFIGURATION: Fluke 5500A ============================================================================= # # Source: # Astro-Med Procedure for ASC-902 # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP {Procedure not formally tested with UUT.} 1.001 DISP {See procedure README file.} 1.002 ASK- R N P F V 1.003 ASK+ K X 1.004 HEAD PRELIMINARY INSTRUCTIONS 1.005 DISP Turn unit ON. 1.005 DISP 1.005 DISP Warm-up time: 30min - REQUIRED 1.005 DISP Temperature: 23 +/-2 degC 1.005 DISP Humidity: 55 +/-10% 1.006 DISP [32] WARNING 1.006 DISP HIGH VOLTAGE is used or exposed during the performance 1.006 DISP of this calibration. DEATH ON CONTACT may result if 1.006 DISP personnel fail to observe safety precautions. 1.007 DISP Extension Card: Astro-Med part number 30776-001 1.008 HEAD SET-UP PROCEDURE 1.009 DISP Turn UUT OFF! 1.009 DISP 1.010 DISP Remove the right side cover of the ASC-902. 1.010 DISP 1.010 DISP Connect the extension card to the signal conditioner PCB 1.010 DISP and reinsert into the card cage. 1.011 DISP Set the ASC-902 switches to the following positions: 1.011 DISP [32] Filter: OFF 1.011 DISP [32] Zero Suppression: OFF 1.011 DISP [32] Mode: P-P 1.011 DISP [32] Coupling: DC 1.011 DISP [32] Polarity: NON-INV 1.012 DISP Place the Negative meter lead to common input 1.012 DISP (black binding post on back of card cage). 1.012 DISP 1.012 DISP Turn UUT ON. 1.013 DISP Short the inputs together by placing a jumper between 1.013 DISP the positive, negative, and common (red, grey, black) 1.013 DISP binding posts on the back of the card cage. 1.014 HEAD {INPUT NULL} 1.015 DISP Connect the 34401A to the UUT as follows: 1.015 DISP [32] 34401A Input HI to TP6 1.015 DISP [32] 34401A Input Lo to signal ground 1.016 DISP Set the Attenuator/Range to 1mV/div. 1.017 34401 0.000V 0.001U 2W 2.001 HEAD {RANGE NULL} 2.002 DISP Set the Attenutor/Range to 1V/div. 2.003 34401 0.000V 0.001U 2W 3.001 HEAD {FINAL NULL} 3.002 DISP Connect the 34401A to the UUT as follows: 3.002 DISP [32] 34401A Input HI to pin 2 of edge connector on the 3.002 DISP [32] BNC connector on the back card cage (signal out) 3.002 DISP [32] 34401A Input Lo to signal ground 3.003 34401 0.000V 0.005U 2W 4.001 HEAD {RMS NULL} 4.002 DISP Set the Attenuator/Range to OFF. 4.002 DISP Set the Mode to RMS. 4.003 34401 0.000V 0.005U 2W 5.001 HEAD {P-P GAIN} 5.002 DISP Connect the 5500A to the UUT binding posts as follows: 5.002 DISP [32] 5500A NORMAL HI to UUT Positive (red) 5.002 DISP [32] 5500A NORMAL LO to UUT Negative (grey) 5.002 DISP 5.002 DISP Leave the 34401A connected to the UUT. 5.003 DISP Set Mode to P-P 5.003 DISP Set Attenuator/Range to 10mV per division. 5.004 5500 0.500V S 2W 5.005 34401 0.500V 0.005U 2W 6.001 HEAD {RMS GAIN} 6.002 DISP Set Mode to RMS 6.002 DISP Set Attenuator/Range to 10mV per division. 6.003 5500 1.000V S 2W 6.004 34401 1.000V 0.005U 2W 7.001 HEAD {ZERO SUPPRESSION} 7.002 DISP Remove the input signal. 7.003 DISP Set the Attenuator/Range to 10mV per division. 7.003 DISP Set Zero Suppression dial to 5. 7.003 DISP Set Zero Suppression switch to +. 7.004 34401 5.000V 0.050U 2W 8.001 DISP Set Zero Suppression switch to -. 8.002 34401 -5.000V 0.050U 2W 9.001 END